Full-series premium instruments. Precision technology for every industry MDR6000 distillation on-line anayzer
MDR6000 distillation on-line anayzer

pplicable Testing Standards

ASTM D86, ISO 3405, IP 123; Cetane Index calculation conforms to ASTM D4737 and ASTM D976 (both Test Procedure A and B)

Cetane Index Supporting Function

Capable of measuring density (API gravity) with automatic cetane index calculation


Operating Parameters

ParameterSpecification
Single test cycle< 10 minutes
Distillation repeatability±1.5°C (accuracy equivalent to or better than ASTM D86)
Measurement accuracyDistillation accuracy superior to D86, IP-123, ISO 3405; Density accuracy ±0.0005 g/mL
Temperature/Distillation rangeFull boiling curve 20–400°C (68–752°F)
Temperature resolution±0.1°C

Sample Requirements

ParameterSpecification
Sample consumption< 2 L/hour (sample must be homogeneous, single-phase, free of free water)
Sample back pressureAtmospheric to < 8 bar (120 psig), must be 1.4 bar (20 psig) below feed pressure
Sample flow rate10–20 L/hour (adjustable based on process conditions)
Sample temperatureWithout cooler: minimum temperature must be 15°C below initial boiling point; With cooler: maximum temperature up to 30°C above initial boiling point (depending on conditions)
Sample feed pressure2–10 bar (30–145 psi)

Cabinet, Installation, and Utility Requirements

ParameterSpecification
Dimensions110 mm (L) × 675 mm (H) × 385 mm (W)
Weight75 kg
Standard ambient operating temperature5–40°C (41–104°F); Optional built-in heater extends range to –10–40°C (14–104°F)
Hazardous area certificationNEC Class 1, Division 2, Groups B/C/D; CSA, ATEX certified Ex II 2 G II T4 Gb (Zone 1); Cabinet heater version: Ex II T3 Gb
Power supply200–240 V AC, max 6A, 50/60 Hz, 700W
Cabinet purge airFlow rate > 150 L/min (5.3 SCFM), pressure 4–10 bar (72–145 psi)
Inert purge nitrogenInlet pressure 3–10 bar (43–145 psi), purity > 95%
Installation site requirementsOutdoor installation requires shade and weatherproof shelter

Terminal Signal Interfaces

ParameterSpecification
Analog outputs4 isolated 4–20 mA signals, user‑configurable to output specified cut point data
Digital I/O8 digital inputs, 8 digital outputs, user‑configurable; Optional 4 additional outputs
Serial communicationSingle bidirectional RS232/RS485, Modbus protocol capable of transmitting full distillation curve; Optional Ethernet TCP/IP

Optional Accessory Modules

ModuleDescription
Density meterIntegrated high‑precision density measurement module
Nitrogen generatorOptional when no external nitrogen supply is available
Sample heating moduleFor samples with cloud point below ambient temperature; ATEX/UL/CSA certified
Sample cooling moduleFor low‑initial‑boiling‑point media such as gasoline; ensures stable sample inlet temperature
Multi‑stream moduleExpandable to up to 2 additional sample streams
IIC flame arrestorSuitable for hydrogen and acetylene applications
External USB keyFor data backup and export; ATEX/CSA certified
Extended analog output boardAdds 4 configurable 4–20 mA analog signals
Mounting kitFloor‑standing rolling stand or wall‑mounting bracket available